logo

Scopri il potenziale illimitato della tua ricerca con i nostri strumenti all'avanguardia.

Contatti Via L. Da Vinci 2C - Besozzo (VA) +39 0332 773210 2biol@2biol.com
#

Description

The carbon dioxide cell culture incubator is an indispensable element in the process of cell culture, finds application in a wide variety of fields and ensures the control of three basic parameters (temperature, CO2 concentration and relative humidity). The built-in 140℃ dry sterilization program helps prevent bacterial contamination. All data and performance trend graphs can be viewed for 7 days on the 7-inch touch screen, and stored data can be downloaded via USB.

Following are some details:

  • High-temperature resistant infrared (IR) CO2 concentration transducer enables real-time monitoring of carbon dioxide level change. To monitor humidity trends in real time, a dedicated probe is available (optional)
  • Heating on all boundary surfaces in conjunction with a Pt1000 platinum thermistor significantly reduces the time to reach the target temperature.
  • The 7-inch touch screen provides complete visibility of data, including the three main parameters (temperature, CO2 and rH).
  • High-temperature sterilization at 140℃ ensures elimination of potential contamination. The built-in HEPA filter enables an ISO 5 (Class 100) quality inner chamber.
  • Display of the parameter curve for the past 7 days and recording of 8000 operations/alarms ensure the safety of samples and the incubator. Stored data can be downloaded via USB.

Technical Specifications

td {border: 1px solid black;}

Temperatures Control & Range ±0.1℃@ 37℃,5℃above ambient to 50℃
Heating Method Air jacket(direct heat)
Uniformity ±0.3℃ @ 37℃
Alarm Setting Customized, Upper limit: (T+0.1) ~ (T+5.0); Lower limit:(T-0.1) ~ (T-5.0)
Sensor Pt1000 Platinum thermistor
Controller PID
Readability & Setability 0.1℃
Sterilization Method 140℃ dry heat sterilization
CO2 Concentration Control & Range ±0.1%@5%,0-20%
Sensor Heat-resistant infrared (IR) sensor
Inlet Pressure 1.0 bar (recommended)
Uniformity ±0.3%
Readability & Setability 0.10%
Alarm Setting Customized, Upper limit: (C+0.1) ~ (C+1.0); Lower limit: (C-0.1) ~ (C-5.0)
Humidity Humidity Pan Volume ~3.8 L
Display 1% RH
Shelves Dimensions 48.7×46.1 cm
Surface Area 224.5 cm2
Number (Standard & Max) 4 ; 17
Size & Weight Interior Volume 188 L
Interior Dimensions 51.0(W)×54.7(D)×67.5(H) cm
Exterior Dimensions 69.5(W)×67.0(D)×97.0(H) cm
Net Weight 110 kg