Technical specifications.
td {border: 1px solid black;}
Accuracy |
±0.5% |
Syringes (Min/Max) |
0.5 µl to 1000 µl |
Flow Rate |
Minimum (0.5 µl syringe): 3.66 pl/min |
Maximum (1000 µl syringe): 3.818 ml/min |
Display |
4.3 WQVGA TFT Color Display with Touchpad |
Non-Volatile Memory |
Stores all Settings |
Connectors |
RS485 – IEEE-1394 6 pos |
USB – Type B |
I/O & TTL – 15 pin D-Sub Connector |
Footswitch – mini phono jack |
Linear Force (Max) |
5 kg (11 lb) @ 100% Force Selection |
Drive Motor |
1.8° Stepper Motor |
Motor Drive Control |
Microprocessor with 1/16 microstepping |
Number of Microsteps |
3,2 |
for one rev.of Lead Screw |
Step Rate – Minimum |
27.5 sec/µstep |
Step Rate – Maximum |
52 usec/µstep |
Pusher Travel Rate – Minimum |
0.433 µm/min |
Pusher Travel Rate – Maximum |
228.97 mm/min |
Input Power |
12-30 VDC |
Input Power Connection |
2.5 mm ID x 5.5 mm OD male plug |
Power |
100-240 VAC: 50/60 Hz, 50 W, 0.5 A fuse |
Dimensions |
22.6 x 17.78 x 9.32 cm (9 x 7 x 3.67 in) |
Weight |
1.96 kg (4.32 lb) |
Operating Temperature |
4 degree C to 40 degree C (40 degree F to 104 degree F) |
Storage Temperature |
-10 degree C to 70 degree C (14 degree F to 158 degree F) |
Humidity |
20% to 80% RH non condensing |
Mode of Operation |
Continuous |
Classification |
Class 1 |
Pollution Degree |
1 |
Installation Category |
11 |
Regulatory Certification |
CE, EU RoHS |