Description

Optogenetics uses light-sensitive proteins to control biological processes. That is, it involves exploiting certain light-sensitive proteins that are appropriately inserted into specific cell populations so that, irradiated by an appropriate and very precise light, they open and the cells become positively or negatively polarized (instantaneous excitation and de-excitation). The unparalleled spatiotemporal accuracy, high tuning, reversibility, and few side effects of light make optogenetics a powerful strategy for studying neuroscience and complex biological systems. For these reasons, optogenetics has already revolutionized disciplines such as neuroscience, developmental biology and cell signaling and is poised to do the same with the study of cell metabolism. Common lines of research include the following: neural circuits and networks, learning and memory, addictions, dyskinesia, sleep disorders, Parkinson’s disease, depression, anxiety, and more.

In its different configurations, this instrument can provide the following wavelengths: 450nm, 465nm, 532nm, 589nm and 630nm (only one wavelength can be selected for each device).

Data and configurations can be saved in the internal memory or in a USB memory.

Technical Specifications

Power:0-100mW(IOS-635 is 1-60mW adjustable),1mW resolution

Stability:<1%

Remote control:10m

Input:2V-10V

Output:10mV-5V

Trigger mode:Edge/Real-Time/Gate

Edge mode-1:Ifthe trigger voltage is higher than or equal to the set value of trigger voltage, the program will run; the trigger is only featured by the “on” function and the program’s duration depends on its set duration.

Real-time mode-2:Ifthe trigger voltage is higher than or equal to the set value of trigger voltage, the program will run; if it is lower than that value, the program will stop; the trigger is featured by the “on/off” functions. The program’s duration depends on the trigger’s duration.

Gate mode-3:therunning power relies on the first subprogram selected and other parameters are defined by the external device and only square wave at the frequency ≤500Hz are responded to.

Safety:password;high temperature alarm;program run prompt

Storage:internal memory/ USB flash disk

Wavelength:450, 465, 532, 589 , 630 and 635nm (One wavelength can be selected per instrument)